Table-top Pulse Oximeter
•High resolution 4.3’’ color LCD screen •Compact and portable, allowing for uninterrupted monitoring •Parameter: SpO2, pulse rate •AC power supply, with built-in lithium battery •Adjustable audible and visual alarms •Storage and review of patient’s trend data up to 36 hours •Applicable to adult, pediatric and neonatal patient •Convenient and quick in inquiring measurement data •USB data upload •Optional: NIBP, ETCO2, printer
Performance
SPO2 monitoring
Continuous real-time measuring. Adopt digital SPO2 technology, which has strong anti-interference capability. Real-time display SPO2, pulse rate, SPO2 plethysmogram, bar graph, SPO2 and Pulse rate (PR) trend graph.
Trend graph with the function of the adaptive scaling.
Standard adult finger-clip probe, binding for neonatal is optional.
SPO2 Measuring
·Measurement range: 35%-100%
·Accuracy: ±2% on the stage of 70%-100%, unspecified in stage of less than 70%.
Pulse Rate Measuring
Measurement: range: 30BPM-240BPM
·Accuracy: ±2BPM or ±2%, whichever is greater.
Measurement Performance in Weak Filling Condition: SPO2 and pulse rate can be shown correctly when pulse-filling rate is 0.4%, SPO2 error is ±4%, the pulse rate error is ±2BPM or ±2%, whichever is greater.
Resistance to surrounding light: the deviation between the value measured in the condition of man-made light or indoor natural light and that of darkroom is less than ±1%.
|